New Products

 

 

New Products

Thromboteck PSe, LupoTek DetecTin VL,

and LupoTek CorrecTin VL

 

See below for more details.

__________________________________________

 

THROMBOTEK PSe

Clotting Assay for Quantification of Protein S Activity

Product # 90-480

 

 

·         Complete kit for the quantitative determination of Protein S activity in human plasma.

 

·         Performed as a simple factor assay.  Suitable for manual and fully automated methods.

 

·         Precision:  4.9% repeatability, 5.7% total imprecision for normal plasma

                        7.8% repeatability, 9.2% total imprecision for abnormal plasma

 

·         Linearity:  10% PS to 156% PS (maximum tested)

 

·         Lower limit of detection:  1% Protein S activity

 

·         Reconstituted stability:  24 hour 2-8°C

                                               8 hour RT

 

·         Typical calibration data (AMAX 200):

 

Calibrator Standard

 (Actual %)

PSe Lot PS0130 clot time, sec

PSe Lot PS0140 clot time, sec

PSE Lot PS0150 clot time, sec

156.00%

131.6

120.2

125.5

117.00%

114.1

103.0

106.9

78.00%

98.9

89.9

92.0

39.00%

81.9

73.5

74.3

19.50%

72.1

64.6

64.3

10.00%

66.3

60.3

59.5

5.00%

63.2

57.0

56.0

 

 

 

 

r squared

0.997

0.998

0.998

slope

44.6

41.0

45.1

intercept

62.7

56.4

55.3

 

 

 

 

Click Here  to View Product Insert

 

 

Back to Main Page

 

__________________________________________________________

 

LupoTek DetecTin VL

    and LupoTek CorrecTin VL

 

r2 Diagnostic's new venom screen and confirm reagents for detection of Lupus Anticoagulants.

 

LupoTek DetecTin VL Product # 85-202

LupoTek CorrecTin VL Product # 86-201

 

·                     LupoTek DetecTin VL and CorrecTin VL use Vipera lebetina (VLV) venom rather than

Vipera russelli (RVV) venom.

 

·                     VLV activates FX to FXa via the same mechanism as RVV.

 

·                     Using VLV is more eco-friendly than using RVV.

 

·                     LupoTek DetecTin VL and CorrecTin VL demonstrated sensitivity of 100% and specificity of

100% with a normalized ratio of 1.30 in a 122 patient cut-off study.

 

·                     LupoTek DetecTin VL and CorrecTin VL demonstrated 98% positive agreement, 96% negative

agreement, and 97% overall agreement with dRVV reagents in a 155 patient, multi-site study.

 

*r2 is a U.S. manufacturer and is the owner of its VLV intellectual property*

 

 

Click Here to View Product Insert

 

 

Back to Main Page