THROMBOTEK PSe

Clotting Assay for Quantification of Protein S Activity

Product # 90-480

 

 

·        Complete kit for the quantitative determination of Protein S activity in human plasma.

 

·        Performed as a simple factor assay.  Suitable for manual and fully automated methods.

 

·        Precision:  4.9% repeatability, 5.7% total imprecision for normal plasma

                        7.8% repeatability, 9.2% total imprecision for abnormal plasma

 

·        Linearity:  10% PS to 156% PS (maximum tested)

 

·        Lower limit of detection:  1% Protein S activity

 

·        Reconstituted stability:  24 hour 2-8°C

                                               8 hour RT

 

·        Typical calibration data (AMAX 200):

                   

 

Calibrator Standard

 (Actual %)

PSe Lot PS0130 clot time, sec

PSe Lot PS0140 clot time, sec

PSE Lot PS0150 clot time, sec

156.00%

131.6

120.2

125.5

117.00%

114.1

103.0

106.9

78.00%

98.9

89.9

92.0

39.00%

81.9

73.5

74.3

19.50%

72.1

64.6

64.3

10.00%

66.3

60.3

59.5

5.00%

63.2

57.0

56.0

 

 

 

 

r squared

0.997

0.998

0.998

slope

44.6

41.0

45.1

intercept

62.7

56.4

55.3

 